Compat Enteral Feeding Pump User Manual
- Compat Ella is an innovative new portable enteral feeding pump with a unique coloured user-friendly interface in local language.
- Compat Enteral Feeding Pump is versatile, lightweight and easy to use feeding pump designed for the administration of enteral tube feeding formulas. Compat Enteral Feeding Pump is versatile, lightweight and easy to use feeding pump designed for the administration of enteral tube feeding formulas.
Compat Enteral Feeding Pumps - English Instructional Video. Get your team aligned with all the tools you need on one secure, reliable video platform.
Kangaroo epump, Enteral feeding pump, Loading & rinsing of feeding pump
1
2
3
4
COVIDIEN, COVIDIEN with logo and ™ marked brands are
trademarks of Covidien AG or an affiliate.
©
2009 Covidien. All rights reserved.
H6212-5M-0409
Proper rinsing of the feeding set
6)
Power the pump on. Select “Keep
Settings” to continue with previous
setting or “Clear Settings” to start
with programming new pump settings.
1)
Feed & flush valve assembly,
2)
Feed only valve assembly,
3)
Distal end of feeding set,
4)
Pump
1)
Grasp thumb tab on valve and
insert firmly into left pocket, ensure
valve is fully seated. Tab should align
with raised white line on left.
2)
Grasp black ring retainer, gently
wrap tubing around rotor, and insert
retainer directly into right pocket.
Do not overstretch tubing.
1)
Disconnect feeding set from patient
before rinsing. The feeding set must
remain loaded during the rinsing
procedure.
2)
CAUTION
– if set is removed from
the pump, it may damage the set and
prevent use and will negatively affect
the system accuracy.
3)
Empty contents of feeding bag.
5)
Empty the feed bag and repeat
the last step until the bag contains
clean water.
Click Run to start the installation immediately. Click Save to copy the download to your computer for installation at a later time. Windows 8.1 manual update download. To start the download, click the Download button and then do one of the following, or select another language from Change Language and then click Change.
10)
CAUTION
– To achieve proper
accuracy, the bottom of the feeding
set bag must be 18' above the top
of the feeding pump.
7)
Press the “Prime Pump” button to
enter the “Prime Pump” screen.
8)
Press and hold the “Hold to Prime”
button until only clean water flows
through the tubing. Once the rotor
starts, this takes between 1 and 2
minutes.
9)
Empty bag of water. Prime to
remove any remaining water out of
the feeding set.
18'
g
4)
Fill the feed bag with water
to approximately half of the total
capacity. Properly seal the bag and
shake for 5 to 10 seconds.
3)
To properly unload a pump set,
power the pump off and wait 10 sec-
onds before removing the feeding set.
15 Hampshire Street
Mansfield, MA
02048
1.800.962.9888
508.261.8000
www.covidien.com
Loading & Rinsing of Feeding Pump
Proper loading & unloading of a feeding set
Do not use feeding sets for longer than 24 hours.
g
Kangaroo ePump
™
Enteral Feeding Pump