Kjeltec 8400 Analyzer Unit User Manual Average ratng: 6,8/10 8062 reviews

Operation

Analyzer for falling number Alphatec and hammer mill Hammertec. The Kjeltec™ 8100 distillation unit provides a simple and reliable solution for safe, semi-automatic distillations. Kjeltec™ 8400 - fully automatic distillation unit. A fully automated analyser, Kjeltec™ 8400, with on. KjeltecTM 8400 Sampler Kjeltec TM 8420 Sampler Kjeltec 8460 Weight net 35 kg 71 kg 89 kg Dimensions (w × d × h) 48 × 58 × 69 cm 64,5 × 84 × 90 cm 100 × 80 × 90 cm Accessories. Compass, PC software for Kjeltec™ 8400 Analyser unit. Tank for titrant, complete with tubing. Burette Dosing device for calibration of burette.

2012 qx56 manual download

  1. Turn power on.
  2. Turn printer on.
  3. Turn water on.
  4. Go to manual mode.
  5. Flick titrating tube to get bubbles out. Dispense titrant several times to get bubbles out.
  6. Go to steam on. Press arrow key to turn steam on. Allow to warm up for 5 minutes.
  7. Press arrow key to turn steam off.
  8. Go to analysis mode.
  9. Select program.
  10. Go to result. Select blank.
  11. Have tube containing water in place.
  12. Close door.
  13. Run at least two blanks. Make sure the blank value is around 0.10 or less.
  14. Go to result and change to % Nitrogen. (The blank number above should read the last blank value you ran.)
  15. Put a tube with 1 mL of the 1% N solution standard and some water in place. Input 1.0 for the weight.
  16. Close door and let run. Readout should be close to 1% (+ 0.05%).
  17. Proceed with samples. Run the glycine check first. Input the correct weight for each sample. (The result for the glycine check should be close to 18.66 % N.)
  18. If you have blanks within the set run a tube with just water first to flush out the system, then run the blank. (Change result to blank for running the blank then back to % N for the samples. Input the weight for the samples as above.)
  19. Record numbers from printout into nitrogen book and onto lab sheets with 2 decimals.

Calculations

Kjeltec
  • Import of kjeltec. Click to View; Customs Import Duty of: kjeltec. Click to View. KJELTEC 8200 SYSTEM NO.91784177 CONSISTING OF BASIC UNIT KJELTEC 8200: SER.NO:91783853 MATERIAL NO.60063212: Denmark. 60063869 - KJELTEC 8400 (PROTEIN ANALYZER MANUFACTURED BY FOSS ANALYTICAL, MOSEL KJELTEC 8400 ANALYZER WITH ACCESSORIES) China: Bombay.
  • Mar 24, 2014  The Kjeltec™ 8200 Auto Distillation unit is the ultimate solution for safe, automatic distillation. Moisture Analyzer. Kjeltec™ 8400/8420/8460 Auto Sampler Systems - Duration.
  • Possibility to upgrade Kjeltec 8200 Distillation Unit to Kjeltec 8400 Analyzer Unit. Automatic dilution, alkali addition, distillation and tube emptying provide ease of use. Variable output steam generator broadens application area to other volatile components. Official and accurate procedures (ISO, AOAC, EPA, DIN) simplify validation.

Foss Kjeltec 8100

  1. %N = 14.01 x N H2SO4 x (mL acid)/(sample wt(g) x 10)
  2. % Crude Protein = %N X 6.25

Shut Down and Maintenance

Daily

  • Place tube with water in unit. Select manual. Go to steam on. Press arrow key and allow generator to run for about 5 minutes. Press arrow key to turn steam off. Remove tube and dispose of contents.
  • Turn power off.
  • Turn water off.
  • Turn printer off.
  • Clean drip tray. Clean the shield. Rinse and inspect the rubber adapter. Wipe any spillage with a damp cloth. Squirt water into titration vessel and onto pins to clean.
  • Place empty tube in unit and close shield.

Weekly

  • Check and clean around reagent tank screw caps to remove crystals. Every one to three months:
  • Clean alkali pump. Put 2 L of warm distilled water in a beaker. Put alkali tube into beaker. Select manual mode. Select add alkali several times. Empty distillation tube as needed. (A rinse using a half a liter of warm water is sufficient.) Replace hose into alkali tank. Flush out residual water and remove entrapped air.
  • Clean receiver solution dispensing system as above only if contamination has occurred.
  • Check volume of alkali. Go to manual mode. Select the pump you want. Measure dispensed liquid. Adjust as necessary.
  • Check volume of receiver solution. Put drain tube into measuring cylinder. Go to manual mode. Select add receiver. Adjust as necessary.
  • Clean splash head. Put 25 mL distilled water and 25 mL acetic acid into distillation tube. Select manual. Go to steam on for 5 to 10 minutes. Turn off steam. Replace tube with tube containing 50 mL distilled water. Distill for 5 minutes. Repeat the distillation three times with fresh water.

Yearly

Kjeltec 8400 Analyzer Unit User Manual
  • Inspect all tubing and connectors.

References

Foss Kjeltec 8400 Manual

  1. Brenner, J. M. 1965. Total Nitrogen. In C. A. Black (ed) Methods of Soil Analysis, Part II, SSSA., Madison, Wis. p.1149-1178.
  2. Horwitz, W. (ed). 1970. Official Methods of Analysis of the AOAC, 11th Edition, Washington, D. C. p.16-17.
  3. Tecator Manual. Kjeltec System 2300 Distilling Unit. Tecator, Hoganas, Sweden.